Description Of The Class "Environment, Display - A&D GP-61KS Instruction Manual

Gp series precision balance
Table of Contents

Advertisement

9-4 Description of the Class "Environment, Display"

Condition (
)
Cond
Cond 0
Cond 2
Stability band width ( 5t-b )
This item controls the width to regard a weight value as a stable value. When the fluctuation per
second is less than the parameter, the balance displays the stabilization indicator and outputs or
stores the data. The parameter influences the "Auto print mode"
5t-b 0
5t-b 2
Hold function ( Hold ) (Animal weighing mode)
This function is used to weigh a moving object such as an animal.
When the weighing data is over the weighing range from zero and the display fluctuation is within the
stabilization range for a fixed period of averaging time, the processing indicator illuminates and the
balance displays the average weight of the animal. When the animal is removed from the weighing
pan, the display returns to zero automatically.
This function is available only when the hold function parameter is set to "1" (the animal mode
indicator ANIMAL illuminates) and any weighing unit other than the counting mode is selected.
The stabilization range and averaging time are set in "Condition (Cond)" and "Stability band width (5t-b)".
Weighing range
GP-12K/20K
GP-30K/30KS/32K/32KS/
40K/61K/61KS
GP-60K/60KS/100K/100KS
GP-102K
This parameter is for sensitive response to the fluctuation of a weight value.
Used for powder target weighing, weighing a very light sample or when
quick response weighing is required.
After setting, the balance displays FAST.
This parameter is for stable weighing with slow response. Used to prevent a
weight value from drifting due to vibration or drafts.
After setting, the balance displays SLOW.
Notes
In automatic response adjustment, the response rate is selected
automatically.
With "Hold function (Hold)" set to "ON (1)", this item is used to set the
averaging time.
This parameter is for sensitive response of the stabilization indicator. Used
for exact weighing.
This parameter ignores slight fluctuation of a weight value. Used to prevent
a weight value from drifting due to vibration or drafts.
Note
With "Hold function (Hold)" set to "ON (1)", this item is used to set the
stabilization range.
Cond 0
Cond 1
20 g
Cond 2
50 g
Averaging time
2 seconds
Faster
4 seconds
8 seconds More accurate
40
Stabilization range
5t-b 0
|
5t-b 1
5t-b 2
Small
|
Big

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents