Agilent Technologies 2000 X-Series Service Manual page 54

Hide thumbs Also See for 2000 X-Series:
Table of Contents

Advertisement

2
Testing Performance
54
Pmeas
= 687 uW
max_freq
Std Dev(n)
max freq
Then after converting the values from the power meter to
Vrms:
response(dB) = 20 log
15 The result from step 14 should be between +3.0 dB and
- 3.0 dB. Record the result in the Performance Test Record
(see
page
63).
16 Move the power splitter from the channel 1 to the channel
2 input.
17 Turn off the current channel and turn on the next
channel using the channel keys.
18 Repeat steps 3 through 17 for the remaining channels,
setting the parameters of the channel being tested where
appropriate.
= 161.6 mV
161.6
mV
/
185.3
 
10
210.4
mV/
211.2
2000/3000 X-Series Oscilloscopes Service Guide
mV
= -1.16 dB
 
mV