Agilent Technologies 2000 X-Series Service Manual page 44

Hide thumbs Also See for 2000 X-Series:
Table of Contents

Advertisement

2
Testing Performance
44
Table 11
Settings Used to Verify Dual Cursor Accuracy, 2000 X-Series
Models
Volts/Div
Oscilloscope
Setting
Calibrator Setting
5 V/Div
35 V
2 V/Div
14 V
1 V/Div
7 V
500 mV/Div
3.5 V
200 mV/Div
1.4 V
100 mV/Div
700 mV
50 mV/Div
350 mV
20 mV/Div
140 mV
10 mV/Div
70 mV
1
5 mV/Div
35 mV
1, 2
2 mV/Div
14 mV
1, 2
1 mV/Div
7 mV
1
A blocking capacitor is required at this range to reduce noise. See
Blocking Capacitor to Reduce Noise"
2
Full scale is defined as 32 mV on the 2 mV/div range and the 1 mV/div range.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
Test Limits
33.6 V
13.44 V
6.72 V
3.36 V
1.344 V
672 mV
336 mV
134.4 mV
67.2 mV
33.2 mV
12.56 mV
5.56 mV
on page 48.
2000/3000 X-Series Oscilloscopes Service Guide
to
36.4 V
to
14.56 V
to
7.28 V
to
3.64 V
to
1.456 V
to
728 mV
to
364 mV
to
145.6 mV
to
72.8 mV
to
36.8 mV
to
15.44 mV
to
8.44 mV
"Use a