2
Testing Performance
Digital
Multimeter
BNC-Bananna
Cable
32
Oscilloscope
Calibrator
BNC Tee
Figure 2
4 Use a BNC- banana cable to connect the multimeter to the
other side of the BNC Tee.
5 Connect the BNC Tee to the Channel 1 output of the
calibrator as shown in
6 On the oscilloscope, press the [Digital] key, then press the
Thresholds softkey, then press the D7 - D0 softkey repeatedly
until the check mark is next to User.
7 Press the User softkey to the right of the D7 - D0 softkey,
then turn the Entry knob (
2000/3000 X-Series Oscilloscope
Test
Connector
Setting Up Equipment for Digital Channel Threshold Accuracy
Test
Figure
2000/3000 X-Series Oscilloscopes Service Guide
Probe
Cables
Channels
8 - 15
Channels
0 - 7
2.
) on the front panel of the