2000/3000 X-Series Oscilloscopes Service Guide
Table 12
Settings Used to Verify Dual Cursor Accuracy, 3000 X-Series
Models
Volts/Div
Oscilloscope
Setting
Calibrator Setting
5 V/Div
35 V
2 V/Div
14 V
1 V/Div
7 V
500 mV/Div
3.5 V
200 mV/Div
1.4 V
100 mV/Div
700 mV
50 mV/Div
350 mV
20 mV/Div
140 mV
10 mV/Div
70 mV
1
5 mV/Div
35 mV
1, 2
2 mV/Div
14 mV
1, 2
1 mV/Div
7 mV
1
A blocking capacitor is required at this range to reduce noise. See
Blocking Capacitor to Reduce Noise"
2
Full scale is defined as 32 mV on the 2 mV/div range and the 1 mV/div range.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
c Press the [Acquire] key.
d Then press the Acq Mode softkey and select Averaging.
e Then press the #Avgs softkey and set it to 64.
Wait a few seconds for the measurement to settle.
4 Press the [Cursors] key, set the Mode softkey to Normal,
then press the X Y softkey and select Y. Press the Y1
softkey, then use the Entry knob (labeled
front panel) to set the Y1 cursor on the baseline of the
signal.
Testing Performance
Test Limits
34.0 V
to
13.6 V
to
6.8 V
to
3.4 V
to
1.36 V
to
680 mV
to
340 mV
to
136 mV
to
68 mV
to
34 mV
to
13.2 mV
to
6.2 mV
to
on page 48.
2
36.0 V
14.4 V
7.2 V
3.6 V
1.44 V
720 mV
360 mV
144 mV
72 mV
36 mV
14.8 mV
7.8 mV
"Use a
on the
45