Agilent Technologies 2000 X-Series Service Manual page 38

Hide thumbs Also See for 2000 X-Series:
Table of Contents

Advertisement

2
Testing Performance
38
Table 9
Settings Used to Verify DC Vertical Gain Accuracy,
3000 X-Series Models
Volts/Div
Oscilloscope
Setting
Calibrator Setting
5 V/Div
35 V
2 V/Div
14 V
1 V/Div
7 V
500 mV/Div
3.5 V
200 mV/Div
1.4 V
100 mV/Div
700 mV
50 mV/Div
350 mV
20 mV/Div
140 mV
10 mV/Div
70 mV
1
5 mV/Div
35 mV
1, 2
2 mV/Div
14 mV
1, 2
1 mV/Div
7 mV
1
A blocking capacitor is required at this range to reduce noise. See
Blocking Capacitor to Reduce Noise"
2
Full scale is defined as 32 mV on the 2 mV/div range and the 1 mV/div range.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
d Press the [Acquire] key.
e Then press the Acq Mode softkey and select Averaging.
f Then press the #Avgs softkey and set it to 64.
Wait a few seconds for the measurement to settle.
4 Add a measurement for the average voltage:
a Press the [Meas] key.
b Press Source; then, turn the Entry knob (labeled
on the front panel) to select the channel you are
testing.
Test Limits
34.2 V
13.68 V
6.84 V
3.42 V
1.368 V
684 mV
342 mV
136.8 mV
68.4 mV
34.2 mV
13.36 mV
6.36 mV
on page 41.
2000/3000 X-Series Oscilloscopes Service Guide
to
35.8 V
to
14.32 V
to
7.16 V
to
3.58 V
to
1.432 V
to
716 mV
to
358 mV
to
143.2 mV
to
71.6 mV
to
35.8 mV
to
14.64 mV
to
7.64 mV
"Use a