8.2.12 Display low cut On/Off

When the low output is cutoff by the setting in 8.2.10 "Low cutoff value," whether the instantaneous
flow rate, flow velocity and % value should be set to zero or not can be selected.
Selection
1: ON
2: OFF
Checking whether the display low cut is On or Off
Proceed as follows to check whether the display low cut is On or Off.
Switch operation
D4: DSP LOW CUT
D4: DSP LOW CUT
Display low cut is effective (On)
Display (instantaneous flow rate, flow velocity and %) will be set to
zero when the low output is cutoff.
Display low cut is not effective (Off)
Measured value will be output even if the low output is cutoff.
Display example
Select D4: DSP LOW CUT from the setting item selection
menu.
2:OFF
The currently set display low cut On/Off status appears.
Returns to the menu of setting item selection.