8.5 Zero Adjustment
8.5.1 Still Water Zero Adjustment
Still water zero adjustment must be performed when the fluid in the measuring pipe of the detector
stays still.
Switch operation
−
Note2
Note2
−
Display example
10.00 m/s
Measured value being displayed (measurement mode)
100.0 %
Pressing
or more, the screen moves to the menu lock cancellation
ENTER111
screen.
000
* Pressing
returns to the measurement mode.
Set the menu lock cancellation cord [111] by
pressing
ENTER111
* Pressing
111
returns to the measurement mode.
Press
>1:SET
mode and then the mode changes to the mode selection screen.
2:CNT CTRL
Select 1:SET in the mode selection screen and then the menu
A1:EX CURRENT
screen appears.
Press
E1: ZERO ADJUST
ZERO ADJUST.
The current flow rate measured value in % unit appears.
E1: ZERO ADJUST
(Check that the fluid to be measured stays still.)
01.0%
ADJUST READY appears and the system is ready for
ADJUST READY
adjustment. (If "CANCEL" is pressed, Adjustment ready
01.0%
condition will be cancelled.)
About 3 to 6 seconds later, the still water zero adjustment ends
E1: ZERO ADJUST
and the measured flow value after adjustment appears.
* ZERO ADJUST
E1: ZERO ADJUST
Returns to the menu of setting item selection.
00.0%
Press
E1: ZERO ADJUST
number becomes 0.
- 118 -
Description
in the measurement mode for 3 seconds
MENU/ENT
in the mode selection screen, the mode
CANCEL
and
.
in the mode selection screen, the mode
CANCEL
for 3 seconds or more in the measurement
MENU/ENT
and
so that the function number becomes E1:
and
so that the number digit of the function
6F8A0917