8.5 Zero Adjustment

8.5.1 Still Water Zero Adjustment

Still water zero adjustment must be performed when the fluid in the measuring pipe of the detector
stays still.
Switch operation
Note2
Note2
Display example
10.00 m/s
Measured value being displayed (measurement mode)
100.0 %
Pressing
or more, the screen moves to the menu lock cancellation
ENTER111
screen.
000
* Pressing
returns to the measurement mode.
Set the menu lock cancellation cord [111] by
pressing
ENTER111
* Pressing
111
returns to the measurement mode.
Press
>1:SET
mode and then the mode changes to the mode selection screen.
2:CNT CTRL
Select 1:SET in the mode selection screen and then the menu
A1:EX CURRENT
screen appears.
Press
E1: ZERO ADJUST
ZERO ADJUST.
The current flow rate measured value in % unit appears.
E1: ZERO ADJUST
(Check that the fluid to be measured stays still.)
01.0%
ADJUST READY appears and the system is ready for
ADJUST READY
adjustment. (If "CANCEL" is pressed, Adjustment ready
01.0%
condition will be cancelled.)
About 3 to 6 seconds later, the still water zero adjustment ends
E1: ZERO ADJUST
and the measured flow value after adjustment appears.
* ZERO ADJUST
E1: ZERO ADJUST
Returns to the menu of setting item selection.
00.0%
Press
E1: ZERO ADJUST
number becomes 0.
- 118 -
Description
in the measurement mode for 3 seconds
MENU/ENT
in the mode selection screen, the mode
CANCEL
and
.
in the mode selection screen, the mode
CANCEL
for 3 seconds or more in the measurement
MENU/ENT
and
so that the function number becomes E1:
and
so that the number digit of the function
6F8A0917

Advertisement

Table of Contents
loading

Table of Contents