Remote Still Water Zero Adjustment

Still water zero adjustment can be performed using an external signal.
To use this function, follow the procedure in 8.2.14, "Digital I/O Function" and set the DI as a zero
adjustment start signal.
[Signal input condition]
Zero adjustment signal
* The signal must be set to H level first and then it must go to L level 10 seconds later but not more
than 20 seconds, and then zero adjustment starts at the falling edge of the signal.
If the signal stays in H level less than 10 seconds or more than 20 seconds, the signal will be
ignored.
Zero
10 seconds
adjustment
start
20 seconds