100% Flow Rate Calibration (Span Calibration) - Toshiba L5232 Instruction Manual

Electromagnetic flowmeter converter
Table of Contents

Advertisement

9.2.3 100% Flow Rate Calibration
Using the Mag-Prover internal calibration circuit, 100% flow rate value (hereafter called the span
value) calibration can be performed.
Checking the span value
Switch operation
P3:FLOW CAL100
P3:FLOW CAL100
Calibrating the span value
Switch operation
P3:FLOW CAL100%
ADJUST READY
P3:FLOW CAL100
* CAL. 100% ADJ.
P3:FLOW SCAL100
P3:FLOW CAL100
Note 1: To cancel the adjustment from the condition of ADJUST READY, press
to the condition where 100% flow rate value using the simulation input is displayed.
Note 2: Calibration time differs depending on the exciting frequency.
Display example
Select P3:FLOW CAL 100 from the calibration item selection
menu.
100.1 %
Span value using the simulation input appears.
Returns to the menu of calibration item selection.
Display example
Select P3:FLOW CAL 100 from the calibration item selection
menu.
100.1 %
Span value using the simulation input appears.
Press
is ready for span calibration. (Note 1)
100.1 %
The display shown left appears and the span calibration starts.
About 3 to 6 seconds later (Note 2), the span calibration ends
and a new span value appears.
100.0 %
Returns to the menu of calibration item selection.
- 125 -
(Span Calibration)
Description
Description
. Then the display shown left appears and the system
6F8A0917
. The display returns
CANCEL