8.2.10 Low Cutoff value

The low cutoff is the fuction to forcefully fix the current output to 0% when the flow rate becomes
equal or below the low cutoff value set near 0%.
The low cutoff value can be set from 0 to 10% and in increments of 0.1%.
If 0% is set, the low cutoff function does not work.
Checking the low cutoff value
Proceed as follows to check the low cutoff value.
Switch operation
D2: LOW CUT
D2: LOW CUT
Display example
Select D2: LOW CUT from the setting item selection menu.
The currently set low cutoff value appears.
01.0 %
Returns to the menu of setting item selection.