Changing the zero offset adjustment value
Changing the zero offset adjustment value from +1.0% to − 2.5%.
Switch operation
O1: MANUAL ZERO
O1: MANUAL ZERO
O1: MANUAL ZERO
O1: MANUAL ZERO
O1: MANUAL ZERO
Note: If you try to set a value outside of the settable range, an error message of *H. OVER SPEC or *L. OVER
SPEC appears. Try to set a new value again.
Display example
Select O1: MANUAL ZERO from the setting item selection menu.
The currently set zero offset adjustment value appears.
+001.0%
The cursor appears.
Press
+001.0%
change.
Change the number of the digit.
-001.0%
To change other digit, press
the cursor to that digit and change the number there.
-002.5%
The display blinks and the set value is displayed for
confirmation.
Press
-002.5%
disappears and the display returns to the condition that the
setting value is displayed.
Returns to the menu of setting item selection.
- 121 -
Description
further to change the cursor to the desired digit to
again to write the data. Then the cursor
MENU/ENT
6F8A0917
under this condition to move