8.2.8 Span Value
The following constants are set here.
(1) Range type
(2) Span unit (changeable only for Range 1)
(3) Span value
(4) Hysteresis
Range type
By selecting the range type, the multi-range function can be used in which the range in operation is
switched between the multiple ranges depending on whether the flow rate is value large or small, or
the flow direction is forward or reverse.
If it is not necessary to use the multi-range function, set the Range type to Single range.
RANGE TYPE
1:SINGLE
2:4F-0R
3:2F-2R
4:EXT.2F-0R
5:EXT.2F-2R
6:EXT.4F-0R
Note: To use the multi-range function, the following DO and DI are required:
In the case of internal automatic switching ranges
2 ranges, forward/reverse range·························1 Digital Output
3 ranges or more, forward/reverse 2 ranges·······2 Digital Outputs
In the case of external signal switching ranges
2 ranges, forward/reverse range·························1 Digital Inputs
3 ranges or more, forward/reverse 2 ranges·······2 Digital Inputs
(Range)
Single range
Single direction, 4 ranges
Forward/reverse direction, 2 ranges
Single direction, 2 ranges
Forward/reverse direction, 2 ranges
Single direction, 4 ranges
- 72 -
RANGE TYPE
Internal automatic switching
Internal automatic switching
External signal switching
External signal switching
External signal switching
6F8A0917