6 Locate the wires and fuse to their original position.
7 Refit the blower module.

Replacing the power supply (VPAP III and VPAP III ST models)

1 With the blower module removed, lift the power supply out of its mounting slots.
2 Disconnect the AC input cable and the DC output cable.
DC output cable
AC input cable
3 Install the new power supply, reconnecting both cables, and ensuring that the power supply is
correctly located in the mounting slots. Make sure the wires are located as shown below:
5 Replacing major subassemblies
29