13 Connect tube 3 to the blower outlet connector on the blower.
14 Ensure that the flexible cable is not stuck underneath the blower.
Note:
If replacing the sensor tubes ensure they are cut to the following lengths:
Slimline
pressure
port fitting
"L" shaped
connector

Replacing the blower inlet and outlet connectors

1 Grasp both halves of the blower module near the blower outlet connector in order to pull it
apart.
5 Replacing major subassemblies
Tube 1
170 ± 5 mm
Tube 2
225 ± 5 mm
Tube 3
120 ± 5 mm
25