10 Take the top half of the blower casing and place it on the lower half of the casing. Ensure the
top half of the casing sits behind the blower outlet connector as shown below:
11 Press evenly around the edge of the casing the reseal the blower casing.

Replacing the fuse (for VPAP III and VPAP III ST models only)

1 With the blower module removed the fuse casing is now accessible. It is located under the
folded PCB insulator.
2 Pull out the wires to remove the fuse casing from behind the insulator.
3 Push the casing together, and twist in an anti-clockwise direction. The fuse casing will separate,
giving access to the fuse.
4 Replace the blown fuse.
5 Reconnect the fuse casing
28
VPAP® III, VPAP III ST, VPAP III ST-A Service Manual
Correct
Incorrect
PCB insulator which
covers the fuse casing