4 Feed the short blue wire in the gap between the brown and blue wires running to the AC
connector, as shown below:
5 Connect the short blue wire to terminals 1a and N.
38
VPAP® III, VPAP III ST, VPAP III ST-A Service Manual
Short blue wire
Short blue wire connected to
terminal 1a and N