ResMed VPAP III, VPAP III ST, VPAP III ST-A Service Manual page 23

Flow generator and humidifier
Hide thumbs Also See for VPAP III, VPAP III ST, VPAP III ST-A:
Table of Contents

Advertisement

3 Turn it over towards the left and rest it on the jig.
4 Using a small screw driver, carefully prize off the two silicone tubes connected to the base of
the blower.
5 Disconnect the motor cable assembly from the Main PCB.
6 Place the new blower module on the jig.
7 The sensor tube that was connected to the blower outlet connector is connected to the third
sensor port on the Main PCB. Ensure that the elbow is facing towards the side of the case as
shown below: