Voltage[:Dc] - Agilent Technologies E1326B User Manual

5 1/2 digit multimeter
Table of Contents

Advertisement

:VOLTage[:DC]

Parameters
Comments
Chapter 5
CONFigure:VOLTage[:DC] [<
DC voltage function and allows you to specify the range and resolution. If you
specify a channel list, those multiplexer channels are scanned.
For a complete listing of range and resolution values available, see
Table 5-1 on page 120.
Parameter Name
Parameter Type
numeric
range
<
>
numeric
resolution
<
>
numeric
channel_list
<
>
The
:DC
parameter is optional. Both of the following command statements select
the DC voltage function:
CONF:VOLT:DC
To select a standard measurement range, specify range as the input signal's
maximum expected voltage. The multimeter then selects the correct range.
The
AUTO
and
DEFault
(enable autorange). The
integration time to 1 PLC.
The
MIN
and
MAX
parameters select the minimum or maximum values for range
and resolution:
For range:
MIN
= 0.113 V;
For resolution:
MIN
selects the best resolution (the smallest value from Table 5-1) for
the selected range.
MAX
for the selected range.
The channel list is of the form (
= card number and
cc
for example).
To select autorange, specify
value for the parameter. In the autorange mode, the multimeter samples the input
signal before each measurement and selects the appropriate range.
To specify a
MIN
or
MAX
CONF:VOLT:DC AUTO
range
resolution
>[,<
Range of Values
0.113 V | 0.91 V | 7.27 V |
58.1 V | 300 V | AUTO | DEF |
resolution
DEF | MIN | MAX
Refer to the multiplexer user's manual for
a list of channels available.
or
CONF:VOLT
options for the range parameter have the same effect
option for the resolution parameter defaults the
DEF
MAX
= 300 V.
selects the worst resolution (the largest value from Table 5-1)
@ccnn)
,
(@ccnn,ccnn)
= channel number (105 is channel 05 of card number 1,
nn
AUTO
(or
DEFault
resolution while autoranging, you must specify
or
CONF:VOLT:DC DEF
channel_list
>]] [,<
>]
Default Units
MIN | MAX
(see Table 5-1) |
, or
(@ccnn:ccnn
) for range or do not specify a
(you cannot omit the range
Command Reference
selects the
volts
volts
none
), where
132

Advertisement

Table of Contents
loading

This manual is also suitable for:

E1411b

Table of Contents