FREQ REFERENCE

Display Outputs

Recorder Outputs

SWEEP

4-6 Characteristics
Connector
BNC female. rear Dane1
50 Q, nominal
Frequency
10 MHz f50 Hz
0 dBm to + 10 dBm
Amplitude
Phase Noise and
Analyzer performance will be degraded unless
Spurious Signals
frequency reference phase-noise and spurious
signals
are 5140 dBc/Hz, referred to 10 MHz at a 100 Hz
to 10 kHz offset.
X, Y, and Z outputs for auxiliary CRT displays. Compatible with
HP 1300 series displays.
Amplitude
X
Y
Z
BLANK
Rise Time
X
Y
Outputs to drive all current HP X-Y recorders using positive pencoils
or TTL penlift input.
and
[
]
[
LOWER LEFT
UPPER RIGHT
and VIDEO outputs with 0,O and 10,l V respectively, for adjusting X-Y
recorders.
A voltage proportional to the horizontal sweep of the frequency
sweep generator.
Connector
Amplitude
BNC female, rear panel
1 V for full deflection
1 V for full deflection
0 to 1 V intensity modulation, -1 V blank
TTL level >2.4 V for blanking.
Compatible with most oscilloscopes.
keys calibrate the recorder SWEEP
]
BNC female, rear panel
0 V for left edge, + 10 V for right edge
1.7 k D nominal