General
HP-IB Interface
Functions
Environmental
Note
Power Requirements
Humidity
X-Rays
3-l 6 Specifications
Temperature
Operation:
0 "C to 55 "C
Storage:
-40 "C to 75 "C
Increased internal temperatures may result if the rear-panel air filters
are not cleaned regularly.
Altitude
Operation:
Storage:
50 to 60 Hz; 100, 120, 220, or 240 volts (+5%, -10%); approximately
450 VA (40 VA in standby). 400 Hz operation is available as Option
400.
Operation
Except as noted in electrical specifications, type tested at <95%
relative humidity, 25 "C to 40 "C for 5 days.
Storage
5% to 90% relative humidity, 0 "C to 40 "C.
Conducted and radiated interference is within the requirements
of MIL-STD 461C, Part 7 RE02 and CEO3 (Air Force), and
within the requirements of CISPR Publication 11 and
Messempfaenger-Postverfuegung 526/527/79 (Kennzeichnung Mit
F-Nummer/Funkschutzzeichen).
IF Section Serial Prefix 3004A and above
X-rays generated by this instrument are sufficiently screened.
Die in diesem geraet entstehende roentgenstrahlung ist ausreichend
abgeschirmt.
accel. voltage / beschl. spg < 20 kV
IF Section Serial Prefix 3001A and below
When operating, this instrument emits x-rays; however, it is well
shielded and meets safety and health requirements of various
countries, such as the X-Ray Radiation Act of Germany.