15. Frequency
Response
Specification
Related Performance
Test
Description
INPUT #l
INPUT #2
fl.O dB, 100 kHz to 1500 MHz (210 dB RF Attenuation)
Frequency Response Test
The spectrum analyzer rear panel 1ST LO OUTPUT is connected to a
tracking generator which supplies the input signal for the analyzer. A
power meter is connected to a power splitter, along with the tracking
generator, which in turn is connected directly to the SIGNAL INPUT
connector of the analyzer. The signal amplitude is measured from
1 MHz to 1500 MHz in 20 MHz increments and compared to the
measured value of the input signal indicated by the power meter.
This procedure is performed for both input connectors. The operator
is asked to make the new connection to the other input connector
mid-way through the test.

15. Frequency Response

Operation Verification 5-25