17. Third Order
Intermodulation
Distortion
Specification
Related Performance
Test
Description
Intercept is greater than +5.0 dBm from 100 kHz to 10 MHz and
greater than + 10 dBm from 10 MHz to 1500 MHz.
Harmonic and Intermodulation Distortion Test
Two signals with 10 kHz and 1 MHz separation are applied to the first
mixer. The frequencies of third order intermodulation products are
calculated and the spectrum analyzer is set to measure the amplitude
of these responses. The third order intercept is calculated from the
measurements for each separation.
17. Third Order Intermodulation Distortion
Operation Verification 5-27