Output Hold; Figure 3-10: Output Hold For Manual Calibration - Teledyne 7600 Operating Instructions Manual

Ndir infrared gas analyzer
Hide thumbs Also See for 7600:
Table of Contents

Advertisement

Infrared Gas Analyzer
To close the "Setting of Auto Calibration" or cancel the mode
midway, press the ESC key. A previous screen will return.
Settings range:
Hold setting:
Response time:
Average period: 1 to 59 min or 1 to 4 hours (Initial value: 1 hour)
Backlight Timer: 1 to 60 min (Initial value: OFF)
Maintenance mode: 0000 to 9999 (Initial value: 0000)
3.5.7.1 O
H
UTPUT
OLD
By setting an output hold to ON, the output signal of each channel is
held during the calibration (manual calibration and auto calibration) plus
the gas flow time as shown in Figures 3-10 through 3-12. Refer to Section
3.5.4, Setting of Auto Calibration for a description of the gas flow time.
Regardless of Hold ON/OFF setting, an output signal can always be held
via an external input.
Manual calibration

Figure 3-10: Output Hold for Manual Calibration

Teledyne Analytical Instruments
0 to 100% FS
1 to 60 sec. (Initial value: 15 sec)
Operation
75