Operation
• Station No.:
• Range setting: Moves to the screen on which measuring
To adjust or make changes to these parameters:
1. Press the UP or DOWN key to move the cursor to the
item whose setting is to be changed.
2. The values for password, oxygen correction, limit, and
station no. are highlighted.
3. Press the UP or DOWN key to change the value and then
press the ENT key.
Note: If you forget the password you will be locked out from the
maintenance mode. Keep a record of the password in a
secure place. Contact the factory for instruction on
reclaiming forgotten or lost passwords.
To set or change the range:
The measuring range can be arbitrarily selected in the minimum
and the maximum range specified at the time of purchase. The range to
be used can be selected as 1 or 2.
1. Move the cursor to the Range Setting option using the UP
or DOWN key, and then press the ENT key.
2. This brings up another screen that lists the channels and
components. Move the cursor to the channel whose
setting is to be changed by pressing the UP or DOWN
key, and then press the ENT key.
Teledyne Analytical Instruments
Set the station No. for MODBUS
communication. Settable in the range from
00 to 32.
range is changed.
Model 7600
88