Infrared Gas Analyzer
3.5.4.5.1 M
E
ANUAL
XECUTION OF
To perform the auto calibration cycle manually:
1. From the measurement mode, press the MODE key to
display the user mode.
2. Point the cursor to "Setting of Auto Calibration" by
pressing the UP or DOWN key and press the ENT key.
3. In the "Setting of Auto Calibration" item selection screen
that appears, point the cursor to "Auto Calibration Run"
by pressing the
key.
4. "Run" is highlighted, displaying a message to confirm the
execution of auto calibration. Press the ENT key to
execute the auto calibration, or press the ESC key to
cancel.
3.5.4.5.2 F
E
ORCIBLY
To abort an auto calibration event:
Teledyne Analytical Instruments
A
C
UTO
ALIBRATION
UP or DOWN
A
C
ND AN
UTO
ALIBRATION
Operation
keys. Press the ENT
63