Teledyne 7600 Operating Instructions Manual page 66

Ndir infrared gas analyzer
Hide thumbs Also See for 7600:
Table of Contents

Advertisement

Operation
In the "Setting of Auto Zero Calibration" screen that
3.
appears, point the cursor to any item you want to set by
pressing the
4. In the "Auto Zero Calibration Parameter Setting" screen
that appears, perform the value entry or the setting. For
the value entry or setting change, use the UP or DOWN
key. To change the setting, use the SIDE key to move the
cursor to the right.
5. After changing the setting, press the ENT key to accept
the change. Auto zero calibration will now be carried out
using the entered setting value.
To close the "Setting of Auto Calibration" or cancel the mode
midway, press the ESC key. A previous screen will return.
The contacts for the auto zero calibration are closed during auto
zero calibration.
Auto Zero Calibration Example
Start Time
Cycle
Flow Time
ON/OFF
Figure 3-8 shows the auto zero calibration with the above settings.
Teledyne Analytical Instruments
UP or DOWN
key. Press the ENT key.
SUN
12:00
12
hour
300
sec
ON
Model 7600
66