Infrared Gas Analyzer
4.4 Optical Zero Adjustment
If the following operation is not performed correctly,
subsequent analysis may adversely be affected. This
operation should be performed by a trained technician or
the unit sent back to the factory for service.
The following adjustment is performed at reassembly after
removing and cleaning the sample cell.
1. Remove the top cover.
2. Allow dry N
gas inlet until the reading stabilizes. The sample gas is
Teledyne Analytical Instruments
Figure 4-6: Cell Block Assembly
or air to flow through the analyzer sample
2