Teledyne 7600 Operating Instructions Manual page 114

Ndir infrared gas analyzer
Hide thumbs Also See for 7600:
Table of Contents

Advertisement

Appendix
Upper/lower limit alarm:
Instrument error contact output:
Calibration error contact output:
Auto calibration status contact outputs:
Pump ON/OFF contact output:
Teledyne Analytical Instruments
Setting is variable within 1 to 99 hours (in
increments of 1 hour) or Setting is variable
within 1 to 40 days (in increments of 1
day)
Gas flow time setting:
The timing for flowing zero gas in auto
zero calibration is set.
Settable 60 to 900 seconds (in increments
of 1 second)
Alarm contact output turns on when mea-
surement value reach to the preset high or
low limit alarm value.
Contacts close when the channel value of
each channel becomes larger than the high
alarm limit value or smaller than the low
alarm limit value.
Contacts close at occurrence of analyzer
error No. 1, 3 or 10.
Contacts close at occurrence of manual or
auto calibration error (any of errors No. 4
to 9).
Contacts close during auto calibration.
During measurement, this contact close.
While calibration gas is flowing, this con-
tact open. This contact is connected in
power supply of pump, and stop the
sample gas while calibration gas flowing.
Model 7600
114