Oxygen Analyzer
A-1 Models 3010
Digital Interface: Full duplex RS-232 communications port.
Appendix
TAC
Packaging: General Purpose Control Unit
• Flush panel mount (Standard).
• Rack mount — Relay rack mounted to
contain either one or two instruments in one
19" relay rack mountable plate (Optional).
Sensor: L-2 trace Micro-Fuel Cell.
Cell Block: BASEEFA Approved Remote Probe
Ranges: Three user definable ranges.
0–10 ppm to 0–250,000 ppm.
Air calibration range 0-25 %.
Autoranging with range ID output.
Alarms: One system-failure alarm-contact to detect
power failure.
Two adjustable concentration threshold alarms
with fully programmable setpoints.
Diagnostics: Start-up or on-demand, comprehensive, self
testing function initiated by keyboard.
Displays: 2 line by 20 alphanumeric, VFD screen, and
one 5 digit LED display.