Appendix; Model 3000Ma Specifications - Teledyne 3000MA Operating Instructions Manual

Percent oxygen analyzer
Hide thumbs Also See for 3000MA:
Table of Contents

Advertisement

Percent Oxygen Analyzer
A-1 Model 3000MA Specifications
90 % Response Time: 10 seconds at 25 °C (77 °F).
Digital Interface: Full duplex RS-232 communications port.
Operating Temperature: 0-45 °C (32-113 °F).

Appendix

Packaging: General Purpose
• Flush panel mount (Standard).
• Rack mount — Relay rack mounted to
contain either one or two instruments in one
19" relay rack mountable plate (Optional).
Sensor: Paramagnetic Sensor
Ranges: Three user definable ranges from 0-1 % to
0-100 %, plus air calibration range of 0-25 %.
Alarms: One system-failure alarm contact to detect
power failure.
Two adjustable concentration threshold alarms
with fully programmable setpoints.
Displays: 2 line by 20 character, alphanumeric, VFD
screen.
One 5 digit LED display.
Power: Universal power supply 100-240 VAC, at 50/
60 Hz.
Accuracy: ±1% of full scale at constant temperature.
±5% of full scale over operating temperature
Teledyne Analytical Instruments
Appendix
A-1