Maintenance; Routine Maintenance; When To Replace A Cell - Teledyne 3010TAC Operating Instructions Manual

Trace oxygen analyzer
Table of Contents

Advertisement

Oxygen Analyzer
4.1
Aside from normal cleaning and checking for leaks at the gas connec-
tions, routine maintenance is limited to replacing Micro-Fuel cells and fuses,
and recalibration.
Self-diagnostic testing of the system and fuse replacement in the Con-
trol Unit are covered in Part I, chapter 5 of this manual. For recalibration, see
Part I, section 4.4 Calibration.
WARNING: SEE WARNINGS ON THE TITLE PAGE OF THIS
4.2
The characteristics of the Micro-Fuel Cell show an almost constant
output throughout its useful life and then fall off sharply towards zero at the
end. Cell failure in the 3010T is usually characterized inability to zero the
instrument down to a satisfactorily low ppm reading. When this occurs, the
3010TAC system alarm trips, and the VFD displays a failure message.
Before replacing the cell:
a. Check your span gas to make sure it is within specifications.
b. Check for leaks downstream from the cell, where oxygen may be
leaking into the system.
If there are no leaks and the span gas is OK, replace the cell.

Maintenance

Routine Maintenance

MANUAL.

When to Replace a Cell

#.#
CELL FAIL/ ZERO HIGH
Teledyne Analytical Instruments
Part II: Analysis Unit
ppm Anlz
Part II:
4-1

Advertisement

Table of Contents
loading

Table of Contents