Infrared gas analyzer communication functions (modbus)
Hide thumbs Also See for 7500:
Table of Contents

Advertisement

4.

SETTING OF COMMUNICATION CONDITION

In order that the master station and instrument can correctly communicate, following settings
are required.
• All communication condition settings of the master station are the same as those of
instruments.
• All instruments connected on a line are set to "Station Nos. (STno)" which are different
from each other. (Any "Station No." is not shared by more than one instrument.)
4.1

Set items

The parameters to be set are shown in the following table. Set them by operating the front
panel keys.
Item
Transmission speed
Data length
Stop bit
Parity setting
Station No.
4.2

Setting operation

Set the station No. on the analyzer maintenance mode display (see the instruction manual).
Value at
delivery
9600bps
Fixed (cannot be changed)
8 bits
Fixed (cannot be changed)
1 bit
Fixed (cannot be changed)
None
Fixed (cannot be changed)
1
Fixed (cannot be changed)
6
Teledyne Analytical Instruments
Setting range
Remarks
Set the same
communication
condition to the
master station and
all slave stations
Set a different value
to each station.