Specifications; Sensors - Fluke RUSKA 2456-LEM User Manual

Laboratory environment monitor
Table of Contents

Advertisement

Sensors

Operating Range:
Performance:
Notes:
(1)
Precision is defined as the combined effects of linearity, repeatability and
hysteresis.
(2)
Expression of accuracy (uncertainty) conforms with the recommendation of the
ISO Guide to the Expression of Uncertainty in Measurement and includes RSS of
precision, stability, temperature effects, and the calibration standard to 2 sigma
(95%). The expanded uncertainty in the standards utilized by RUSKA to calibrate
the LEM is typically 0.1
Storage Environment:
-20 - 70 ºC; 0 - 95% relative humidity, non-condensing
Electrical Power:
10 - 30 Volt DC input; 250mA max
Temperature:
Pressure:
Humidity:
Precision
Temperature:
Pressure:
Humidity:
Total Uncertainty:
Temperature:
Pressure:
Humidity:
o
C, 2.4% RH, 1.4e-3 kPa.
Chapter 8

Specifications

o
18-28
C
690-1070 mbar (20.4 - 31.6 inHg)
5-95% RH
o
+/-0.1
C
+/-1.4 mbar (+/-0.04 inHg)
+/-3% RH
o
+/-0.5
C/year
+/-2.7 mbar (+/-0.08 inHg)/year
+/-10% RH/year
8-1