The Module number, in the left hand column, is referenced to each LEM as defined in
Figures 2-3 and 2-62-6. To begin logging data for a particular LEM, select the Enable
button.

Intervals

LEM data can be logged into a Day, Week, or Monthly file.
Timing and data information are based upon the clock of the host computer.
Day:
Weekly:
Monthly:

File Names

The file names are generated automatically using the following formats:
Day:
Week:
Month:

Default Location

By default, the log files are placed in
C:\Program Files \ RUSKA Instruments \ LEM Cal \ LEMDataLog \ ...
To change this location, select <Browse Dir> button. Figure 2-9.

File Verification

From the Data Logging screen (Figure 2-9) select Verify Log.
Figure 2-9. Data Logging
Readings are taken at 1 minute intervals.
The active file is saved and a new file is crated at midnight of each day.
Readings are taken at 5 minute intervals.
The active file is saved and a new file is created at midnight of each
Saturday.
Readings are taken at 15 minute intervals.
The active file is saved and a new file is created at midnight of the last
day of each month.
SerialNumber_Year_Day of Year.LOG
Example: SN054323_Y2002_D137.LOG
SerialNumber_Year_Week of Year.LOG
Example: SN054323_Y2002_W47.LOG
SerialNumber_Year_Month of Year.LOG
Example: SN054323_Y2002_M05.LOG