Ending Calibration

When you are using the manual method of calibration and have completed the calibration
of the desired functions (dc volts, ac volts, ohms, and frequency), send CAL_MODE
OFF to clear CAL from the primary display.

RS-232 Device Configuration Parameters

When CAL is shown in the 2680A/2686A display, it indicates that you are in the
calibration mode. The device configuration parameters are set as listed in Table 5-1.
Channel Configuration
Reading Rate
Trigger Type
Intervals
Drift Correction
External Trigger Output
Temperature Units
Queue Overflow Mode
Totalizer Debounce
Autodisable Scanning
Universal
Input
Module
Channel 11
Wires
5-6
Table 5-1. RS-232 Device Configuration for Calibration Procedures
Parameter
11
12
13 14 15 16 17 18 19 20
SOURCE
H L
H L
H L H L H L H L H L H L H L H L
(4-WIRE)
SENSE
H L
H L
H L H L H L H L H L H L H L H L
(4-WIRE)
1
2
3
4
Channel 1
Wires
Figure 5-3. Two-Wire Calibration Connection
Unique to calibration step
Slow
All disabled
All disabled
Enabled
Disabled
Celsius
Discard old
Disabled
On
5
6
7
8
9
10
Setting
5700A Calibrator
OUTPUT
SENSE
V A
V
WIDEBAND
HI
HI
LO
LO
HI
AUX
GUARD GROUND
CURRENT
f4-07.eps