Fluke 2680A Service Manual page 152

Data acquisition system/data logging system
Hide thumbs Also See for 2680A:
Table of Contents

Advertisement

268XA
Service Manual
Front Panel Err
Second 32-bit self-test code
32
33
34
35
36
37
38
39
40
41
42
43
44
45
46
47
48
49
50
51
52
53
54
55
56
57
58
59
6-6
Table 6-1. Self-Test Error Codes (cont)
Self-test
Code
0x00000000
No self-test error.
0x00000001
Module 3 Calibration constants corrupt
0x00000002
Module 3 Calibration Procedures Incomplete
0x00000004
Module 3 A/D failure
0x00000008
Module 3 Zero offset measurements unreasonable
0x00000010
Module 3 Reference balance measurements unreasonable
0x00000020
Module 3 Overload detection failed
0x00000040
Module 3 Open thermocouple detection failed
0x00000080
Module 4 Calibration constants corrupt
0x00000100
Module 4 Calibration Procedures Incomplete
0x00000200
Module 4 A/D failure
0x00000400
Module 4 Zero offset measurements unreasonable
0x00000800
Module 4 Reference balance measurements unreasonable
0x00001000
Module 4 Overload detection failed
0x00002000
Module 4 Open thermocouple detection failed
0x00004000
Module 5 Calibration constants corrupt
0x00008000
Module 5 Calibration Procedures Incomplete
0x00010000
Module 5 A/D failure
0x00020000
Module 5 Zero offset measurements unreasonable
0x00040000
Module 5 Reference balance measurements unreasonable
0x00080000
Module 5 Overload detection failed
0x00100000
Module 5 Open thermocouple detection failed
0x00200000
Module 6 Calibration constants corrupt
0x00400000
Module 6 Calibration Procedures Incomplete
0x00800000
Module 6 A/D failure
0x01000000
Module 6 Zero offset measurements unreasonable
0x02000000
Module 6 Reference balance measurements unreasonable
0x04000000
Module 6 Overload detection failed
0x08000000
Module 6 Open thermocouple detection failed