Fluke 2680A Service Manual page 177

Data acquisition system/data logging system
Hide thumbs Also See for 2680A:
Table of Contents

Advertisement

Table 6-9. Relating Selftest Errors to Device Problems (cont)
Error
Error Code
Code
Description
15
RAM constants
corrupt.
16
Ethernet chip
failure.
17
Module 2
Calibration
constants corrupt.
18
Module 2
Procedures
Incomplete.
19
Module 2 A/D
failure.
20
Module 2 zero
offsets test failed.
21
Module 2
reference balance
test failed.
22
Module 2
overload
detection failed.
Suspect
Assembly
Sys/Ctrl
Background The SYS/CTRL stores constants in flash
PCA
memory, U20, which are transferred to RAM for use in
operation.
Failure For this error to occur, the constants may be
improperly stored in the SYS/CTRL RAM or flash.
Correction This might happen if there is a problem with
SYS/CTRL power supply, RAMs, or the SYS/CTRL U20
flash chip itself.
Sys/Ctrl
Background The Ethernet controller (SYS/CTRL
PCA
U17) supports 10/100BASET transceiver operation.
Failure For this error to occur, the self-test that
simulates Ethernet activity has to fail.
Correction Check Ethernet operation using both
10/100BASE-T. If the Ethernet interface operates over
10BASE-T but not 100 Base-T, then the Ethernet
transceiver SYS/CTRL T1is suspect. If Ethernet does
not operate over 10 or 100Base-T interface, then
SYS/CTRL U17 is suspect. Also, check SYS/CTRL U20
for I/O and memory decoding. See "Troubleshooting the
Digital Kernel" for more information.
Module
Background Same as for module 1.
PCA
Failure Same as for module 1.
Correction Same as for module 1.
Module
Background Same as for module 1.
PCA
Failure Same as for module 1.
Correction Same as for module 1.
A/D
Background Same as for module 1.
Converter
Failure Same as for module 1.
and DIO
PCA
Correction Same as for module 1.
A3 A/D
Background Same as for module 1.
Converter
Failure Same as for module 1.
PCA
Correction Same as for module 1.
A3 A/D
Background Same as for module 1.
Converter
Failure Same as for module 1.
PCA
Correction Same as for module 1.
A3 A/D
Background Same as for module 1.
Converter
Failure Same as for module 1.
PCA
Correction Same as for module 1.
Diagnostic Testing and Troubleshooting
Troubleshooting the Device
Error Code
Discussion
6
6-31