If the PC card has a configuration in the scan data that is inconsistent with the current
module mix, scanning will not begin. When using a PC card, you must either format the
PC card or modify the module mix to match the PC card scan configuration before
scanning can begin.
Self-Test Code
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
3-4
Table 3-1. Self-Test Codes
No self-test errors
Boot block ROM checksum failed
Bad outguard main image in ROM
Configuration data file corrupt
Display test failure
Display not responding
Module 1 calibration constants corrupt
Module 1 calibration procedures incomplete
Module 1 A/D failure
Module 1 zero offset measurements unreasonable
Module 1 reference balance measurements unreasonable
Module 1 overload detection failed
Module 1 open thermocouple detection failed
Communication parameter constants corrupt
Ethernet address corrupt
RAM constants corrupt
Ethernet chip failure
Module 2 calibration constants corrupt
Module 2 calibration procedures incomplete
Module 2 A/D failure
Module 2 zero offset measurement unreasonable
Module 2 reference balance measurements unreasonable
Module 2 overload detection failed
Module 2 open thermocouple detection failed
No module self-test errors
Module 3 calibration constants corrupt
Module 3 calibration procedures incomplete
Module 3 A/D failure