Manual Addenda; Specifications - Fluke 35040 Operator's Manual

Therapy dosimeter
Table of Contents

Advertisement

CAUTION
The
heading used in this manual explains hazards that could damage the instrument.
Such damage may invalidate the warranty.

1.5 Manual Addenda

Any improvements or changes concerning the instrument or manual will be explained in an addendum
included with the manual. Be sure to note these changes and incorporate them into the manual.

1.6 Specifications

Charge Measurements
Range, Resolution and Repeatability:
Full Scale Range
Display Resolution
200.00 pC
0.01
2.0000 nC
0.0001 nC
20.000 nC
0.001 nC
200.00 nC
0.01
2.0000 μC
0.0001 μC
20.000 μC
0.001 μC
200.00 μC
0.01
2.0000 mC
0.0001 mC
20.000 mC
0.001 mC
Measurement Method: Charge flowing into the input is integrated using a capacitive feedback
electrometer whose output is measured by means of a 20 bit bipolar A-D converter. A high-speed reset
circuit removes the integrated charge on the feedback capacitor each time it reaches a level of 2 nC in
either polarity. By monitoring both the number of 2 nC resets and the charge on the feedback capacitor,
the total charge flowing into the input is determined.
Factory Calibration: Basic factory charge calibration is performed by injecting -1.90 nC and
a precision standard air capacitor. This method is believed to provide a calibration accuracy of
0.05%rdg (± 0.05% of reading) as detailed in the table below:
Factory Calibration Error Source
Charge Source Accuracy (Capacitor)
Charge Source Accuracy (Voltage Change)
Temperature Effects (21°C – 25°C)
Other
Maximum Error
The field adjustable current and charge scale factors are set to their nominal value of 1.000 during factory
recalibration.
Repeatability (±
counts)
pC
5
1.5
2.5
nC
1
1
1
μC
1
1
1
± (%rdg)
0.01
0.01
0.02
0.01
0.05