Fluke 35040 Operator's Manual page 61

Therapy dosimeter
Table of Contents

Advertisement

Explanation: The low battery annunciator is a reminder that while running on the battery, there is a
finite amount of energy available. For the battery in this instrument and the energy
consumption of the instrument, when the battery voltage drops below 5.9 volts there is
less than 30 minutes of operating time left.
Symptom:
"Internal Reset Error" displayed.
ADC
Internal Reset Error
Press RESET/MEASURE
and Repeat Exposure
Cause:
Start of exposure detected during reset process.
Action:
Stop the exposure. Press the RESET/MEASURE button. Wait for the "System Not RDY"
annunciator to extinguish. Start the exposure.
Explanation: The instrument measured a current above the start threshold while resetting the
electrometer. It is assumed that the exposure was started while the reset was occurring.
If the exposure did start during the reset, some unknown amount of charge was lost. The
lost charge would lead to an incorrect reading.
Symptom:
"Configuration Error" displayed.
ADC
Configuration Error
Error Number : 16
Cause:
The instrument has lost the calibration or customization data.
Action:
Convert the configuration error number to a binary value using a program such as the
Windows® Calculator and use the table below to determine the extent of the problem.
For example, a configuration error of "18" corresponds to a binary value of "10010" (bits 5
and 2 set), which indicates that the customization for both the ion chamber and settings
are invalid.
Bit
1
2
3
4
5
6
Service Required
Description
Operator settings are
invalid.
Customization settings are
invalid.
Electrometer calibration
settings are invalid.
Bias calibration settings are
invalid.
Ion chamber customization
settings are invalid.
Ion chambers cannot be
read.
Troubleshooting and Service
Troubleshooting
Action
Select the proper settings
from the front panel.
Reinstall settings using the
Customization software.
Do NOT make
measurements. Return
instrument for repair.
Do NOT make
measurements. Return
instrument for repair.
Reinstall ion chambers using
the Customization software.
Return instrument for repair
to correct problem.
4
4-3