Fluke TRACKER 90100 Operator's Manual page 40

Radiation measurement system
Table of Contents

Advertisement

Model 90100
Operators Manual
Low Battery Indication at the Display
WARNING
The Model 35360A TRACKER Display and 35300A
Detector contain no internal operator serviceable
Disassembly of either unit could result in electrical
h k
Always turn the 35360A TRACKER Display's power off
prior to connecting or disconnecting the interface
bl Failure to do so may result in an electrical
h k
When powering the 35360A TRACKER Display from
an AC line, always use a 3-wire grounding type line cord
h as the one supplied with the 90100 TRACKER System.
In addition, only connect the 35360A TRACKER Display to
AC power outlet employing a third wire safety
d
Figure 4-2
Low Battery Flowchart
4-4
Low Battery annuciator is displayed on
35360A TRACKER Display's VFD.
Yes
Operating from
AC power source?
No
Low battery
annuciator displayed?
Commence operation from AC line
power or charge batteries.
Check AC line fuses at rear panel
of 35360A TRACKER Display.
No
Fuses OK?
Replace 35360A TRACKER
Display's AC line fuses.
Refer to Section 5.3.
No
Connect 35360A TRACKER
Display to an AC power source
and press the POWER key.
Yes
Verify voltage present at AC line
receptical by connecting and
powering some other device.
Yes
AC line voltage present?
Yes
35360A TRACKER Display is faulty and
will require repair or replacement.
No