35360A TRACKER Display will not turn on (display will not light)
NOTE
When attempting to turn the instrument on,
firmly press the POWER key and hold
for approximately one second. Always
allow several seconds between retries.
WARNING
The Model 35360A TRACKER Display and 35300A
Detector contain no internal operator serviceable
Disassembly of either unit could result in electrical
h k
Always turn the 35360A TRACKER Display's power off
prior to connecting or disconnecting the interface
bl Failure to do so may result in an electrical
h k
When powering the 35360A TRACKER Display from
an AC line, always use a 3-wire grounding type line cord
h as the one supplied with the 90100 TRACKER System.
In addition, only connect the 35360A TRACKER Display to
AC power outlet employing a third wire safety
d
Replace 35360A TRACKER
Display's AC line fuses.
Refer to Section 5.3.
Figure 4-1.
No Display Flowchart
Connect 35360A TRACKER
Display to an AC power source
and press the POWER key.
Yes
35360A TRACKER Display
VFD illuminates?
Commence operation from AC line
power or charge batteries.
Check AC line fuses at rear panel
of 35360A TRACKER Display.
No
Fuses OK?
Maintenance & Service
35360A TRACKER Display's
VFD does not illuminate when
POWER key is pressed.
No
Operating from AC power source?
No
Verify voltage present at AC line
receptical by connecting and
powering some other device.
Yes
AC line voltage present?
Yes
Return 35360A TRACKER Display
to Keithley for repair.
4
Troubleshooting
Yes
No
4-3