Figure 8-7. Fault 5, Open Heater - Emerson Oxymitter 4000 Instruction Manual

Oxygen transmitter
Table of Contents

Advertisement

Oxymitter 4000
DIAGNOSTIC
ALARMS
CALIBRATION RECOMMENDED
TEST
POINTS
INC
INC
HIGH
LOW
GAS
GAS
DEC
DEC
KEYPAD
Alarms
O2 Heater Open

Figure 8-7. Fault 5, Open Heater

Rosemount Analytical Inc.
A Division of Emerson Process Management
HEATER T/C
HEATER
SW2
O2 CELL
CALIBRATION
O2 CELL mV +
TP1
O2 CELL mv -
TP2
HEATER T/C +
TP3
HEATER T/C -
TP4
CAL
TEST GAS +
TP5
PROCESS -
TP6
% O2
LOI
37260023
e. Fault 5, Open Heater
Figure 8-7 shows the electronic assembly
for an Oxymitter 4000 with a membrane
keypad (upper view) and an Oxymitter 4000
with an LOI (lower view).
Membrane Keypad. When Fault 5 is de-
tected, the HEATER LED flashes once,
J1
pauses for three seconds, and repeats.
1. Remove power.
2. Remove the electronic assembly per
3. Using a multimeter, measure across
4. The measurement should be approxi-
LOI. When Fault 5 is detected, the LOI dis-
plays the "O2 Heater Open" message.
1. Remove power. Unscrew and remove
2. Perform the diagnostic steps 2 through
Instruction Manual
IB-106-340 Rev. 3.0
December 2003
paragraph 9-4d, Electronic Assembly
Replacement.
the heater connector J8.
mately 72 ohms. If the heater is open,
see paragraph 9-4h, Heater Strut
Replacement.
the LOI module from the electronic as-
sembly.
4 shown for the membrane keypad.
Troubleshooting
8-9