Oxygen analyzer with hps heater power supply field module (for use with existing signal conditioning electronics)
Hide thumbs Also See for World Class 3000:
Table of Contents

Advertisement

World Class 3000
GAS FLOW
DIRECTION
VEE
DEFLECTOR
APEX
FILTER

Figure 2-2. Orienting the Optional Vee Deflector

Rosemount Analytical Inc.
A Division of Emerson Process Management
DIFFUSION
ELEMENT
SETSCREW
VEE
DEFLECTOR
P0017

Figure 2-3. Outline of Heater Power Supply

Instruction Bulletin
c. Service Required
1. Power input: 100, 115 or 220 Vac sin-
gle phase, 50 to 60 Hz, 3 amp
minimum. (See label.)
2. Compressed air: 10 psig (68.95 kPag)
minimum, 225 psig (1551.38 kPag)
maximum at 2 scfh (56.6 L/hr) maxi-
mum; supplied by one of the following
(less than 40 parts-per-million total hy-
drocarbons). Regulator outlet pressure
should be set at 5 psi (35 kPa).
(a) Instrument air - clean, dry.
(b) Bottled standard air with step-down
regulator.
(c) Bottled compressed gas mixture
(20.95% oxygen in nitrogen).
(d) Other equivalent clean, dry, oil-free
air supply.
106-300NE Rev. 3.4
May 2000
Installation
2-7