Oxymitter 4000
CALIBRATION RECOMMENDED
alarm signals an IMPS 4000 or SPS
4000 to initiate a calibration. The se-
quencer sends an "in cal" signal to the
control room so that any automatic
control loops can be placed in manual.
Then, the sequencer begins to se-
quence the calibration gases.
3. Manual calibrations must be performed
at the Oxymitter 4000 site and require
operator intervention throughout the
process. Manual calibration instruc-
tions, in condensed form, are on the
inside of the right electronics housing
cover (Figure 9-2). For units with a
membrane keypad, use the following to
perform a manual calibration. For units
with an LOI, refer to paragraph 9-3.
(a) Place control loop in manual.
(b) Verify the calibration gas parame-
ters are correct (paragraph 9-2a).
Rosemount Analytical Inc.
A Division of Emerson Process Management
ALARMS
LED
FLASHES
STATUS
1
OPEN
2
SHORTED
HEATER T/C
3
REVERSED
A/D COMM
4
ERROR
1
OPEN
HIGH HIGH
2
TEMP
HIGH CASE
3
HEATER
TEMP
4
LOW TEMP
5
HIGH TEMP
1
OPEN
3
BAD
O2 CELL
EPROM
4
CORRUPT
1
INVALID SLOPE
CALIBRATION
INVALID
2
CONSTANT
LAST CAL
3
FAILED
SW2 DIP SWITCH
HART
0-10%
4mA
NOT USED

Figure 9-2. Inside Right Cover

MANUAL
CALIBRATION
*
PLACE CONTROL LOOP
IN MANUAL
*
IF CAL LED ON
GO TO STEP 2
PUSH CAL
1
CAL LED ON
PUSH CAL
2
CAL LED FLASH
3
APPLY TG1
PUSH CAL
4
CAL LED ON SOLID
WAIT FOR FLASH
5
REMOVE TG1 & APPLY TG2
PUSH CAL
CAL LED ON SOLID
WAIT FOR FLASH
6
2 FLASH-VALID CAL
3 FLASH-INVALID CAL
7
REMOVE TG2
PUSH CAL
CAL LED ON FOR
8
PURGE TIME
CAL LED OFF
LOCAL
0-25%
20mA
NOT USED
(c) If performing a manual calibration
with CALIBRATION RECOM-
MENDED LED off and the CAL
LED off, start at step 1.
(d) If performing a manual calibration
with CALIBRATION RECOM-
MENDED LED on and the CAL
LED on, start at step 2.
1
2
Instruction Manual
IB-106-340 Rev. 3.0
December 2003
29770005
Push the CAL key. The CALI-
BRATION RECOMMENDED
LED will come on and the
CAL LED will be on solid. If a
multimeter is attached across
TP5 and TP6, the reading will
display the percentage of
oxygen seen by the cell.
Push the CAL key. The CALI-
BRATION RECOMMENDED
LED will turn off and the
CAL LED will flash continu-
ously. The Oxymitter 4000
can be configured so that the
Maintenance and Service