Figure 8-1. O; Figure 8-15. Fault 13, Invalid Slope - Emerson Oxymitter 4000 Instruction Manual

Oxygen transmitter
Table of Contents

Advertisement

Oxymitter 4000
DIAGNOSTIC
ALARMS
CALIBRATION RECOMMENDED
TEST
POINTS
INC
INC
HIGH
LOW
GAS
GAS
DEC
DEC
KEYPAD
Alarms
O2 Cell Bad

Figure 8-15. Fault 13, Invalid Slope

Rosemount Analytical Inc.
A Division of Emerson Process Management
HEATER T/C
HEATER
SW2
O2 CELL
CALIBRATION
J1
O2 CELL mV +
TP1
O2 CELL mv -
TP2
HEATER T/C +
TP3
HEATER T/C -
TP4
CAL
TEST GAS +
TP5
PROCESS -
TP6
% O2
LOI
37260031
m. Fault 13, Invalid Slope
Figure 8-15 shows the electronic assembly
for an Oxymitter 4000 with a membrane
keypad (upper view) and an Oxymitter 4000
with an LOI (lower view).
Membrane Keypad. When Fault 13 is de-
tected, the CALIBRATION LED flashes
once, pauses for three seconds, and re-
peats.
1. During a calibration, the electronics
calculates a slope value. If the value of
the slope is less than 35 mV/deg or
more than 52 mV/deg, the slope alarm
will be active until the end of the purge
cycle.
2. See paragraph 9-2, Calibration. Verify
the calibration by carefully repeating it.
Ensure the calibration gases match the
calibration gas parameters. If you at-
tach a multimeter to TP1+ and TP2-,
sample gas measurements are:
≈ 23 mV
8% O
2
≈ 85 mV
0.4% O
2
3. Power down the Oxymitter 4000 and
remove it from the stack.
4. Replace the cell per paragraph 9-4i,
Cell Replacement.
LOI. When Fault 13 is detected, the LOI
displays the "O2 Cell Bad" message. Refer
to the comments and procedures in para-
graphs 1 through 4 above.
Instruction Manual
IB-106-340 Rev. 3.0
December 2003
Troubleshooting
8-17

Advertisement

Table of Contents
loading

Table of Contents