CONFIGURATION OF OXYMITTER 4000 WITH LOI

Install all protective equipment covers
and safety ground leads before
equipment startup. Failure to install
covers and ground leads could result
in serious injury or death.
4-1

GENERAL

a. Verify Mechanical Installation
Ensure the Oxymitter 4000 is installed cor-
rectly (Section 2, INSTALLATION).
b. Verify Terminal Block Wiring
1. Remove screw (32, Figure 9-3 or
Figure 9-4), gasket (33), and cover lock
(34) that secure the housing cover
(27). Remove the cover to expose the
terminal block (25).
4-20 mA
SIGNAL
LOGIC I/O
GROUND LUGS
Rosemount Analytical Inc.
A Division of Emerson Process Management
SECTION 4
OXYMITTER 4000
SERIAL NO.
TAG NO.
VOLTS:
85-264 VAC 48-62 Hz
OUTPUT:
TERMINAL
BLOCK

Figure 4-1. Electronics Housing Terminals and LOI

2. Check the terminal block wiring, Figure
4-1. Be sure the power, 4-20 mA sig-
nal, and logic outputs are properly
connected and secure.
3. Install the housing cover (27, Figure
9-3 or Figure 9-4) on the terminal block
and secure with cover lock (34), gasket
(33), and screw (32).
4. For an Oxymitter 4000 with an inte-
grally mounted SPS 4000, remove
screws (26, Figure 9-13) and terminal
cover (27). Check that the power and
signal terminations are properly con-
nected to terminal strip (25) and secure
according to instructions in Section 2,
INSTALLATION.
5. Install terminal cover (27) and secure
with screws (26). Make sure terminal
cover gasket (28) is in place.
OXYMITTER 4000
ELECTRONICS
HOUSING
Rosemount Analytical Inc.
SMART FAMILY
R
Orrville, OH 44667-0901
HART
TM
800-433-6076
TM
WATTS:
500 VA
4-20 mA
LINE
FUSE:
5 Amps
Instruction Manual
IB-106-340 Rev. 3.0
December 2003
LOI
37260011
Configuration with LOI