CONFIGURATION OF OXYMITTER 4000
Install all protective equipment covers
and safety ground leads before
equipment startup. Failure to install
covers and ground leads could result
in serious injury or death.
3-1

GENERAL

a. Verify Mechanical Installation
Ensure the Oxymitter 4000 is installed cor-
rectly (Section 2, INSTALLATION).
b. Verify Terminal Block Wiring
1. Remove screw (32, Figure 9-3 or
Figure 9-4), gasket (33), and cover lock
(34) that secure the housing cover
(27). Remove the cover to expose the
terminal block (25).
4-20 mA
SIGNAL
LOGIC I/O
GROUND LUGS

Figure 3-1. Electronics Housing Terminals and Membrane Keypad

Rosemount Analytical Inc.
A Division of Emerson Process Management
SECTION 3
WITH MEMBRANE KEYPAD
Rosemount Analytical Inc.
Orrville, OH 44667-0901
800-433-6076
TM
OXYMITTER 4000
SERIAL NO.
TAG NO.
VOLTS:
85-264 VAC 48-62 Hz
OUTPUT:
4-20 mA
LINE
TERMINAL
BLOCK
2. Check the terminal block wiring, Figure
3-1. Be sure the power, 4-20 mA sig-
nal, and logic outputs are properly
connected and secure.
3. Install the housing cover (27, Figure
9-3 or Figure 9-4) on the terminal block
(25) and secure with cover lock (34),
gasket (33), and screw (32).
4. For an Oxymitter 4000 with an inte-
grally mounted SPS 4000, remove
screws (26, Figure 9-13) and terminal
cover (27). Check that the power and
signal terminations are properly con-
nected to terminal strip (25) and secure
according to instructions in Section 2,
INSTALLATION.
5. Install terminal cover (27) and secure
with screws (26). Make sure terminal
cover gasket (28) is in place.
OXYMITTER 4000
ELECTRONICS
HOUSING
R
SMART FAMILY
HART
TM
WATTS:
500 VA
FUSE:
5 Amps
Instruction Manual
IB-106-340 Rev. 3.0
December 2003
HEATER T/C
DIAGNOSTIC
HEATER
SW2
ALARMS
02 CELL
CALIBRATION
CALIBRATION RECOMMENDED
J1
02 CELL mV +
TP1
TEST
02 CELL mv -
TP2
POINTS
HEATER T/C +
TP3
HEATER T/C -
TP4
INC
INC
CAL
HIGH
LOW
GAS
GAS
TEST GAS +
TP5
DEC
DEC
PROCESS -
TP6
% 02
37260009