Grounding

Step 5
Grounding
The sensor can be grounded via the piping if the joints in the pipeline are ground-bonded. If the sensor is not
grounded via the piping, connect a ground wire to the internal or external grounding screw, which is located
on the transmitter, core processor, or junction box.
Improper grounding can cause measurement error. Ground the flowmeter to earth, or follow ground
network requirements for the facility.
If national standards are not in effect, follow these guidelines to ground the sensor:
• Use copper wire, 14 AWG (2,0 mm²) or larger wire size for grounding.
• Keep all ground leads as short as possible, less than 1 ohm impedance.
• Connect ground leads directly to earth, or follow plant standards.

Pressure Ratings

Table 4 shows the pressure ratings for ELITE sensors.
Table 4
ELITE pressure ratings
EN-1092 flange
PN
Type
Facing Description
40
11
B1
weld
neck
40
11
B2 & D
weld
neck
100
11
B2 & D
weld
neck
40
32/02
B1
lap
joint
1. Pressure rating limited by flange (for 304 sensor tubes).
2. Pressure rating limited by flange (for 316 sensor tubes).
3. Pressure rating limited by flange (for C-22 Alloy sensor tubes).
20
ASTM
material
Flange, EN
304
1092-1 PN40
Form B1 (304)
Flange, EN
316
1092-1 PN40
Form B2 & D
(316)
Flange, EN
316
1092-1 PN100
Form B2 & D
(316)
Flange, EN
C-22 collar/
1092-1 PN40
304 flange
Form B1(lap
joint)
Pressure rating (bar)
≤50°C 100°C 150°C 200°C 250°C 300°C 350°C 400°C 427°C
40.0
34.4
30.8
28.0
40.0
40.0
36.3
33.7
100.0 100.0 90.9
84.2
40.0
34.4
30.8
28.0
26.0
24.1
23.0
22.0
31.8
29.7
28.5
27.4
79.5
74.2
71.4
68.5
26.0
24.1
23.0
22.0
®